User manual EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER

DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!

If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER. We hope that this EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER user guide will be useful to you.


EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER : Download the complete user guide (80 Ko)

Manual abstract: user guide EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER

Detailed instructions for use are in the User's Guide.

[. . . ] EUTECH INSTRUMENTS 4 Installation 2-Wire pH or ORP Transmitter Operating Instructions Operating with one or several Transmitters: Figure 1 - Operating with 1 Transmitter 1 Preface Thank you for purchasing the Eutech pH or ORP Transmitter. [. . . ] Tel: (65) 778 6876; Fax: (65) 773 0836; e-mail: marketing@eutechinst. com; Home page: http://www. eutechinst. com Figure 2 - Operating with several transmitters 5 Electrical Connections 2 Introduction The Transmitter is used for the continuous measurement of pH, ORP (REDOX), Conductivity, Dissolved Oxygen or Temperature, in conjunction with an appropriate sensor. The Transmitter is housed in a IP65 enclosure, with openings for input and 4-20mA output. NOTE: In electrically "noisy" environments, select `Symmetrical' mode of operation and connect a stainless steel rod (Potential Matching Pin - PMP) as shown above. Ensure both the PMP and pH or ORP electrodes are immersed into the solution, even when performing calibration. 3 Diagram and Dimensions (in mm) 6 Before Calibration Remove cover of Transmitter and ensure that electrode is correctly connected as above. Connect Amp meter to the 4/20mA output of Transmitter (Appendix 1). 7 1) 2) pH Transmitter Calibration Calibrate in pH 7. 00 buffer first, and then in either pH 4 or pH 10. First, immerse clean pH electrode into a container with fresh pH 7. 0 buffer solution; 68X256901 Rev 1 10/99 3) 4) 5) 6) 7) 8) 9) Amp meter should read 12mA; if not, calibration is necessary; Locate "Offset adjustment" trimmer; Using a fine screwdriver, slowly turn trimmer till Amp meter reads 12mA; For greater accuracy, adjust slope with pH 4 or 10; Rinse probe in de-ionised water and immerse in either pH 4. 01 or pH 10. 01 buffer solution; In pH 4. 01, the output is 8. 54mA; In pH 10. 01, the output is 15. 44mA; Using a fine screwdriver, slowly turn trimmer, marked `slope', till Amp meter reads correct mA value with respective buffer solutions. Figure 4 - Method Two 11 pH Ordering Information Item Cat Code ECPHXTR ECARTSO05B ECARGTSO05B 10) Remember to disconnect Amp meter and close cover. NOTE: For other buffer values, expected current output is based on: mA = {(16/14) x buffer pH value} + 4mA pH Transmitter 8 1) 2) 3) 4) 5) 6) 7) 8) 9) ORP Transmitter Calibration Prepare fresh Standard solutions, such as 255mV, 470mV. Short ORP inputs and observe Amp meter display; Amp meter should read 4mA; if not, calibration is necessary; Locate "Offset adjustment" trimmer; Using a fine screwdriver, slowly turn trimmer till Amp meter reads 4mA; For greater accuracy, adjust the slope with Standard calibrating solutions; Rinse probe in de-ionised water and immerse in standard solution; In 255mV solution, the output is 8. 08mA; In 470mV solution, the output is 11. 52mA; Using a fine screwdriver, slowly turn trimmer, marked `slope', till Amp meter reads correct mA value with respective Standard solutions. Combination pH Electrode, General, integral 5m cable Combination pH Electrode, w/PMP, integral 5m cable Dissolved Oxygen Item Dissolved Oxygen Transmitter General Purpose D. O. [. . . ] for 24V Item 120V Power Supply 220V Power Supply Cat Code EC-120-XTR EC-220-XTR Housing 77 x 77 x 28 mm, field mountable 10 Appendix 1 Connecting Amp Meter for Calibration Figure 3 - Method One 68X256901 Rev 1 10/99 [. . . ]

DISCLAIMER TO DOWNLOAD THE USER GUIDE EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER




Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual EUTECH INSTRUMENTS 2-WIRE PHORP TRANSMITTER will begin.

 

Copyright © 2015 - manualRetreiver - All Rights Reserved.
Designated trademarks and brands are the property of their respective owners.