If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual EUTECH INSTRUMENTS WATERPROOF CYBERSCAN PC 300 PHCONDUCTIVITYTDS METER. We hope that this EUTECH INSTRUMENTS WATERPROOF CYBERSCAN PC 300 PHCONDUCTIVITYTDS METER user guide will be useful to you.
Manual abstract: user guide EUTECH INSTRUMENTS WATERPROOF CYBERSCAN PC 300 PHCONDUCTIVITYTDS METER
Detailed instructions for use are in the User's Guide.
[. . . ] Instruction Manual
PC 300
Waterproof Hand-held pH/Conductivity/TDS/Temperature Meter
35631-00
Technology Made Easy . . .
68X248908 Rev3 01/04
Preface
This manual serves to explain the use of the Waterproof PC 300 hand-held meter. It functions in two ways, firstly as a step by step guide to help you to operate the meter. It is written to cover as many anticipated applications of the Waterproof PC 300 hand-held meter as possible. If there are doubts in the use of the PC 300, please do not hesitate to contact the nearest Eutech Instruments/ Oakton Instruments Authorized Distributor. [. . . ] To re-select the Auto-ranging function, repeatedly press the RANGE key until the "MEAS" indicator appears without blinking.
Figure 22 - Manual ranging
NOTES If the value of the solution you are measuring is higher than the range selected "Or" will appear on the primary display. You will have reset the manual ranging function each time you turn the meter off.
Figure 23 - Out-of-range
27
Instruction Manual
PC 300
6
HOLD FUNCTION
This feature lets you freeze the display for a delayed observation. HOLD can be used any time in MEAS mode.
1.
To hold a measurement, press the HOLD key while in measurement mode. "HOLD" will appear on the display.
2.
To release the held value, press the HOLD again. Continue to take measurements.
Figure 24 - HOLD function
NOTE: · · · This meter shuts off automatically after 20 minutes of last key press. If the meter is shut off either automatically or manually, the HOLD value will be lost. When this feature is switched on, and when a reading is stable for more than 5 seconds, the display will automatically "HOLD" the reading. To switch on or off the auto endpoint feature, see SETUP program P3. 1 in Section 7. 4 for pH and SETUP, program P7. 1 in Section 7. 8 for conductivity/TDS.
28
Instruction Manual
PC 300
7
ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customized your meter's preferences and defaults. This Waterproof handheld meter features different sub-groups that organize setup parameters. The sub-groups are: For pH P 1. 0 P 2. 0 P 3. 0 P 4. 0 View calibration data (CAL) View electrode data (ELE) Unit Configuration (COF) Reset to factory default P 5. 0 P 6. 0 P 7. 0 P 8. 0 For Conductivity & TDS View calibration data (CAL) View electrode data (ELE) Unit Configuration (COF) Temperature (tPr)
settings (rSt) P 9. 0 Reset to factory default
settings (rSt)
See Figure 25 on next page.
29
Instruction Manual
PC 300
Figure 25 - Overall view of SETUP programs for pH (on left) and conductivity/TDS (on right)
30
Instruction Manual
PC 300
7. 1
1.
Advanced SETUP Mode Overview
In either pH or Conductivity measurement mode, press the SETUP key to enter Set up mode.
2. 3.
Press the
or
keys to scroll through sub groups.
Press ENTER key to enter a particular parameter.
See Addendum 4 for a table of meter factory default settings.
SETUP in pH Measurement Mode
P1. 0: View previous calibration data P1. 1 P1. 2 P1. 3 P1. 4 P1. 5 First calibration point (pH 1. 68) Second calibration point (pH 4. 01) Third calibration point (pH 7. 00 or 6. 86) Fourth calibration point (pH 10. 01 or 9. 18) Fifth calibration point (pH 12. 45)
Instructions on Section 7. 2. P2. 0: View electrode data P2. 1 P2. 2 pH electrode offset pH electrode slope
Instructions in Section 7. 3. P3. 0: Unit configuration P3. 1 P3. 2 P3. 3 P3. 4 READY indicator and auto endpoint function select on or off Number of pH calibration points: 2, 3, 4 or 5 Select buffer group: USA or NIST Select °C or °F
Instructions in Section 7. 4. P4. 0: Reset to factory defaults P4. 0 Reset meter to factory defaults
Instructions in Section 7. 5.
31
Instruction Manual
PC 300
SETUP in Conductivity Measurement Mode
P5. 0: View previous calibration data P5. 1 P5. 2 P5. 3 P5. 4 P5. 5 First range calibration point (0. 00 19. 99 µS) Second range calibration point (0. 0 199. 9 µS) Third range calibration point (0 1999 µS) Fourth range calibration point (0. 00 19. 99 mS) Fifth range calibration point (0. 0 199. 9 mS)
Instructions in Section 7. 6. P6. 0: View electrode data P6. 1 Effective cell constant for first range (0. 00 19. 99 µS) P6. 2 Effective cell constant for second range (0. 0 199. 9 µS) P6. 3 Effective cell constant for third range (0 1999 µS) P6. 4 Effective cell constant for fourth range (0. 00 19. 99 mS) P6. 5 Effective cell constant for fifth range (0. 0 199. 9 mS) Instructions in Section 7. 7. P7. 0: Unit configuration READY indicator and auto endpoint function select on or off P7. 2 Select °C or °F P7. 3 Select Automatic or Manual Temperature Compensation P7. 4 Setting TDS conversion factor Instructions in Section 7. 8. P8. 0: Temperature P8. 1 P8. 2 Adjusting temperature coefficient Adjusting normalization temperature P7. 1
Instructions in Section 7. 9. P9. 0: Reset to factory defaults P9. 0 Reset meter to factory defaults
Instructions in Section 7. 9. 1.
32
Instruction Manual
PC 300
SETUP in TDS Measurement Mode
P5. 0: View previous calibration data P5. 1 P5. 2 P5. 3 P5. 4 P5. 5 First range calibration point (0. 00 9. 99 ppm) Second range calibration point (0. 0 99. 9 ppm) Third range calibration point (0 999 ppm) Fourth range calibration point (0. 00 9. 99 ppt) Fifth range calibration point (0. 0 200 ppt)
Instructions in Section 7. 6. P6. 0: View electrode data P6. 1 Effective cell constant for first range (0. 00 9. 99 ppm) P6. 2 Effective cell constant for second range (0. 0 99. 9 ppm) P6. 3 Effective cell constant for third range (0 999 ppm) P6. 4 Effective cell constant for fourth range (0. 00 9. 99 ppt) P6. 5 Effective cell constant for fifth range (0. 0 200 ppt) Instructions in Section 7. 7. P7. 0: Unit configuration P7. 1 READY indicator and auto endpoint function select on or off P7. 2 Select °C or °F P7. 3 Select Automatic or Manual Temperature Compensation P7. 4 Setting TDS conversion factor Instructions in Section 7. 8. P8. 0: Temperature P8. 1 P8. 2 Adjusting temperature coefficient Adjusting normalization temperature
Instructions in Section 7. 9. P9. 0: Reset to factory defaults P9. 0 Reset meter to factory defaults
Instructions in Section 7. 9. 1.
33
Instruction Manual
PC 300
7. 2
P1. 0: Viewing previous pH calibration data
This mode lets you recall previous pH calibration data, which helps you know when to recalibrate your meter. [. . . ] Ideally T1 and T2 should bracket your measurement temperature, and should not different by more than 5 °C.
4. 5.
Calculate the temperature coefficient of your solution according to the formula shown above. From measurement mode, press the SETUP key to enter Setup mode.
60
Instruction Manual
PC 300
B. F.
Press the
key until the lower display reads P8. 0.
Press the ENTER key twice. [. . . ]