If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual FLUKE MULTIMETER WITH TEMPERATURE 16. We hope that this FLUKE MULTIMETER WITH TEMPERATURE 16 user guide will be useful to you.
Manual abstract: user guide FLUKE MULTIMETER WITH TEMPERATURE 16
Detailed instructions for use are in the User's Guide.
[. . . ] · Do not use the V·Check mode to measure voltages in circuits that could be damaged by this mode's low input impedance (2 k). · Turn off power to the circuit under test before cutting, desoldering, or breaking the circuit. · Do not apply more than 600V rms between a meter terminal and earth ground. [. . . ] Do not use V·Check to measure voltage in circuits that could be damaged by a 2 k load. Hint: V·Check can be effectively used to eliminate "ghost" voltages.
16
MULTIMETER
If AC detected If DC detected Meter puts 2 k load on circuit.
Disable and re-enable V·Check For , and , Disables V·Check and locks meter in selected mode. 2 sec
OFF
A
CAT
V·Check
600V
TEMPERATURE
COM
+
or Re-enables V·Check
or
or
ip09i. eps
Temperature
16
MULTIMETER
F
C
OFF
A
CAT
V·Check
600V
TEMPERATURE
COM
+
80BK Integrated Temperature Probe (Type K)
Vent or Pipe
ip19f. eps
Note correct connector polarity.
To meet stated accuracy, the 80BK temperature adapter must be at the same temperature as the meter.
W Warning
To avoid possible electric shock, DO NOT apply thermocouple tip to any conductor that is greater than 30V AC, 42. 4V pk, or 60V DC to earth.
Capacitance E
Turn off circuit power; then disconnect and discharge the capacitor before measuring capacitance.
16
MULTIMETER
If the capacitor requires more discharging, diSC is displayed while the capacitor discharges.
OFF
A
CAT
V·Check
600V
TEMPERATURE
COM
+
1 µF
10 µF
100 µF
10, 000 µF
+
ip05i. eps
Note correct probe polarity for polarized capacitors.
Microamps µA
Range 0 to 200 µA
16
MULTIMETER
µA AC
µA DC
OFF
A
CAT
V·Check
600V
TEMPERATURE
COM
+
Flame sensor probe
Control module
ip20i. eps
Example: Flame rectification circuit. Turn heating unit on and record µA measurement.
Diode Test G
Turn off circuit power before testing. Also refer to V·Check.
Good Diode
16
MULTIMETER
Typical voltage for silicon diode
16
MULTIMETER
Low Impedance V·Check
OFF
A
CAT
V·Check
OFF
A
CAT
V·Check
600V
TEMPERATURE
600V
TEMPERATURE
COM
+
COM
+
Single beep
ip03i. eps
Forward Bias Bad Diode
16
MULTIMETER
Reverse Bias
16
MULTIMETER
OFF
A
CAT
V·Check
OFF
A
CAT
V·Check
600V
TEMPERATURE
600V
TEMPERATURE
COM
+
COM
+
or
ip04i. eps
Shorted
Open
MIN MAX
µF
(Records the lowest and highest measurements)
V·Check, autoranging, and Battery Saver are disabled. When the reading changes more than about 50 digits, the meter gives a short beep. When a new minimum or maximum is recorded, the meter gives a long beep.
Enter Present Reading with MIN MAX Input Change Alert TM
Maximum Reading
2 sec
To Exit Minimum Reading
ip10i. eps
MIN MAX with Elapsed Time
µF
Records the hours and minutes between when MIN MAX was entered and the last high and low was recorded. To enable the MIN MAX timer, hold down M while turning the rotary switch from OFF to either measurement mode.
Enter MIN MAX. View Present Reading
Maximum Reading
Maximum Reading Elapsed Time
2 sec
Minimum Reading Elapsed Time
Minimum Reading
Exit MIN MAX
ip11i. eps
To disable the beeper for all modes, hold down r for 2 seconds while turning the meter on.
Disabling the Beeper
Continuity Capture
To set up the meter to capture intermittent shorts and opens, turn the switch to GR, connect the leads to the circuit; then press M.
ip12i. eps
Captures transitions longer than 250 µs (1/4000th of a second). Transitions after the first transition cause the meter to beep, but the display does not change. To exit, press M for 2 seconds, or change the measurement mode.
Maintenance
Clean the case with a damp cloth and detergent. Battery Replacement Remove the test leads before disassembling the case.
Replacement Parts Fluke TL-75 (Double-insulated leads) PN 855705 80BK (Integrated Temperature Probe) PN 1273124 Service and Parts This meter should be serviced only by a qualified service technician. To locate an authorized service center, call: USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500 Or, visit Fluke's Web site at www. fluke. com.
C O M
60 0V
+
+
ip13i. eps
Specifications
Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with relative humidity to 90%. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave input. Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits])
Function Temperature Range -10°C to 400°C 14°F to 752°F Resolution 0. 1°C or 0. 2°F 0. 1°C or 0. 2°F Accuracy ±(1. 0% + 0. 8°C) typical ±(1. 0% + 1. 5°F) typical ±(5. 0% + 1. 5°C) typical ±(5. 0% + 3. 3°F) typical Accuracy
(Type K Thermocouple) -40°C to -10°C -40°F to 14°F
Error does not include Type K Thermocouple errors. Function Range 4000 mV 4. 000V 40. 00V 400. 0V 600V 1 4000 mV 4. 000V 40. 00V 400. 0V 600V 400. 0 4. 000 k 40. 00 k 400. 0 k 4. 000 M 40. 00 M 1. 000 µF 10. 00 µF 100. 0 µF 10000 µF
1
Resolution 1 mV 0. 001V 00. 01V 000. 1V 1V 1 mV 0. 001V 00. 01V 000. 1V 1V 0. 1 0. 001 k 0. 01 k 0. 1 k 0. 001 M 0. 01 M 0. 001 µF 0. 01 µF 0. 1 µF 1 µF
±(1. 9% + 3) ±(1. 9% + 3) f ±(1. 9% + 3) (50 to 400 Hz) ±(1. 9% + 3) ±(1. 9% + 3) ±(0. 9% + 2) ±(0. 9% + 2) ±(0. 9% + 1) E ±(0. 9% + 1) ±(0. 9% + 1) ±(0. 9% + 2) ±(0. 9% + 1) ±(0. 9% + 1) ±(0. 9% + 1) ±(0. 9% + 1) ±(1. 5% + 3) ±(1. 9% + 2) ±(1. 9% + 2) ±(1. 9% + 2) | 1000 µF ±(1. 9% + 2) >1000 µF ±(10% + 90) typical ML 2. 000V 0. 001V ±(1. 9% + 2)2 1. [. . . ] This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke's opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption. [. . . ]